7 Reprocessing steps in accordance with ISO 17664-1 /
ISO 17664-2
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Patients, users or third parties could be infected by contaminated medical
devices.
4
Take suitable personal protective measures.
4
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
4
Reprocess the medical device right after treatment.
4
Remove all residual cement, composite or blood immediately.
4
Wipe-disinfect the medical device before transport.
4
Take screw-in and/or snap-on polishing attachments off the medical device.
4
Do not place in solutions or similar substances.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.
KaVo recommends the following products based on the compatibility of the ma-
terials. The microbiological efficacy must be ensured by the disinfectant manu-
facturer and proven by an expert opinion.
▪ CaviWipes and CaviCide made by Metrex
▪ Mikrozid AF made by Schülke & Mayr (Liquid or wipes)
▪ FD 322 made by Dürr
Consumables required:
▪ Cloths for wiping the medical device.
4
Spray the disinfectant on a cloth, then wipe down the medical device and al-
low the disinfectant to act according to the instructions of the disinfectant
manufacturer.
4
Follow the instructions for use of the disinfectant.
7.2 Manual Reprocessing
WARNING
Incomplete disinfection.
Patients, users or third parties could be infected by contaminated medical
devices.
4
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
4
If the disinfectants/disinfection procedures fail to meet the mandatory na-
tional requirements, perform a final sterilisation with the sterilisation para-
meters as described.
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess in a washer disinfector or by hand only.