5 Operation | 5.1 Attaching the medical device

13 / 28

5 Operation

 WARNING

The patient swallows or aspirates loose parts or substances.

Danger of suffocation.

4

Place a rubber dam before each treatment.

NOTE

At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.

5.1 Attaching the medical device

 WARNING

Detaching the medical device during treatment.

A medical device that is not properly locked can detach from the coupling dur-
ing treatment.

4

Before each use, check if the medical device is securely locked onto the
coupling.

CAUTION

Removing and attaching the medical device while the drive motor is ro-
tating.

Damage to the catch pin.

4

Never attach or remove the medical device while the drive motor is rotating.

CAUTION

Pressing the foot switch while attaching or detaching the medical
device.

Property damage to the medical device.

4

Do not connect or remove the medical device while pressing the foot switch.

4

Place the medical device on the motor coupling and lock it into place.

4

Pull on the medical device to make sure that it is securely affixed to the
coupling.

5.2 Removing the medical device

4

Unlock the medical device from the motor coupling by twisting it slightly and
then pulling it along its axis.

5.3 Inserting the dental bur

NOTE

The head housing is protected against the ingress of polishing paste by spe-
cial seals.

NOTE

Only use prophy cups and polishing brushes or supports with Snap-On func-
tion in accordance with DIN EN ISO 13295, support type 5.

Summary of Contents for EXPERTmatic E31 C

Page 1: ...Instructions for use EXPERTmatic E31 C 1 008 4998...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...g for malfunctions and troubleshooting 16 6 1 Checking for malfunctions 16 6 2 Troubleshooting 16 6 2 1 Replacing the O rings on the motor coupling 16 7 Reprocessing steps in accordance with ISO 17664...

Page 4: ...Vo Technical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use ar...

Page 5: ...storage conditions temperature range Transportation and storage conditions air pressure Transportation and storage conditions Humidity Protect from moisture Keep dry Protect from impact Original langu...

Page 6: ...ns for use EXPERTmatic E31 C 1 008 4998 1 User instructions 6 28 CAUTION In cases which if not prevented can lead to minor or moderate injury CAUTION In cases which if not prevented can lead to materi...

Page 7: ...Reprocess the product and accessories appropriately before disposal 4 Use gloves or a finger guard when you test insert and remove the dental bur 2 2 Improper use Because the operation with an electri...

Page 8: ...sories that are equipped with standardised interfaces 4 Do not make any modifications to the device unless these have been ap proved by the manufacturer of the product 4 Use original KaVo spare parts...

Page 9: ...r third parties This may res ult in aspiration swallowing of parts and possibly even a risk of suffocation 4 Only use spare parts that comply with the specification for repair original manufacturer sp...

Page 10: ...r the described application by a properly trained user You need to comply with the following the applicable health and safety regulations the applicable accident prevention regulations these Instructi...

Page 11: ...speed via the motor control 3 3 Transportation and storage conditions CAUTION Startup after refrigerated storage Malfunction 4 Prior to startup strongly refrigerated products must be allowed to warm u...

Page 12: ...uct and accessories before disposal Also refer to 7 Reprocessing steps in accordance with ISO 17664 1 ISO 17664 2 Page 18 Current packaging law Dispose of and recycle the packaging appropriately in ac...

Page 13: ...securely locked onto the coupling CAUTION Removing and attaching the medical device while the drive motor is ro tating Damage to the catch pin 4 Never attach or remove the medical device while the dr...

Page 14: ...ird parties could be infected by contaminated medical devices 4 Use gloves or a finger guard when you test insert and remove the dental bur CAUTION Dental bur with damaged worn or deformed shafts Mate...

Page 15: ...c E31 C 1 008 4998 5 Operation 5 4 Removing the dental bur 15 28 4 Unscrew it or take it off the insert after it has stopped rotating NOTE For the re use of the inserts please refer to the instruction...

Page 16: ...HOUT using KaVo original spare parts Parts could come loose and cause injury Aspiration swallowing of parts and danger of suffocation 4 Only use spare parts that comply with the specification for repa...

Page 17: ...for use EXPERTmatic E31 C 1 008 4998 6 Checking for malfunctions and troubleshooting 6 2 Troubleshooting 17 28 4 Spray a lint free cloth with KaVo Spray apply it to the new O rings and in sert them in...

Page 18: ...only KaVo recommends the following products based on the compatibility of the ma terials The microbiological efficacy must be ensured by the disinfectant manu facturer and proven by an expert opinion...

Page 19: ...internal and external disinfection Manual cleaning Accessories required Tap water drinking water Brush e g medium hard toothbrush 4 Brush off under flowing tap water at least 10 seconds Cleaning with...

Page 20: ...only CAUTION Never reprocess this medical device in an ultrasonic cleaner Malfunction and material damage 4 Reprocess in a washer disinfector or by hand only 7 3 1 pre cleaning Accessories required T...

Page 21: ...ng with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation but no later than after 30 minute...

Page 22: ...F Sterilisation parameters Select a suitable process from the following sterilisation processes depending on the available steriliser Steriliser with triple pre vacuum at least 3 minutes at 134 o C 1...

Page 23: ...ry Mat no Holder 1 004 0596 INTRA Instrument stand 3 005 5204 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 KaVo CLEANspray 2110 P 1 007 0579 KaVo DRYspray 2117 P 1 007 0580 KaVo Spr...

Page 24: ...liable for defects and consequences thereof that have arisen or may arise from natural wear improper handling cleaning servicing or main tenance non compliance with operating maintenance or connectio...

Page 25: ......

Page 26: ......

Page 27: ......

Page 28: ...1 008 5577 bd 20240417 03 en...

Reviews: