10 Information concerning the electromagnetic compatibility according to EN IEC 60601-1-2 | 10.3 Immunity to
electromagnetic interference
174 / 178
Interference im-
munity tests
IEC 60601 test levels
Compliance level
Electromagnetic en-
vironment -
Guidelines
Fast transient electrical
interference / bursts ac-
cording to IEC
61000-4-4
± 2 kV for power lines
± 1 kV for input and out-
put lines
± 2 kV for power lines
The quality of the supply
voltage should corres-
pond to that of a typical
business or
hospital environment.
Surges according to IEC
61000-4-5
± 1 kV voltage phase
conductor - phase con-
ductor
± 2 kV voltage phase
conductor - earth
± 1 kV voltage phase
conductor - phase con-
ductor
± 2 kV voltage phase
conductor - earth
The quality of the supply
voltage should corres-
pond to that of a typical
business or
hospital environment.
Voltage interruptions,
short-term interruptions,
and fluctuations of the
supply voltage according
to IEC 61000-4-11
< 5 % U
T
(> 95 % interruption)
for 1/2 period
40 % U
T
(60 % interruption)
for 5 periods
70 % U
T
(30 % interruption)
for 25 periods
< 5 % U
T
(> 95 % interruption)
for 5 s
(250 periods)
< 5 % U
T
(> 95 % interruption)
for 1/2 period
40 % U
T
(60 % interruption)
for 5 periods
70 % U
T
(30 % interruption)
for 25 periods
< 5 % U
T
(> 95 % interruption)
for 5 s
(250 periods)
The quality of the supply
voltage should corres-
pond to that of a typical
business or hospital en-
vironment. When the
user of this product
needs continued opera-
tion even when the elec-
trical power supply is in-
terrupted, it is recom-
mended to supply this
product from an uninter-
rupted power supply or a
battery.
Magnetic field at a sup-
ply frequency (50/60 Hz)
according to IEC
61000-4-8
3 A/m
3 A/m
Magnetic fields at the
mains frequency should
correspond to typical
values in a business and
hospital environment.
Note: U
T
is the alternating mains voltage before the test level is used.
10.3 Immunity to electromagnetic interference
This product is designed for use in an environment like the one cited below. The
customer or the user should ensure that the device is used in an environment
of the specified type.
Summary of Contents for ESTETICA E70 Vision