Care instructions ESTETICA E70 Vision / E80 Vision
3 Replacing and checking cleaning and germ reduction agents | 3.3 Check the OXYGENAL 6 concentration
The touchscreen shows a status message when the bottle is taken out of the
unit. In order to display the "No Oxygenal bottle" status message, tap first the
arrow and then the status message. Then the required steps can be done.
▶
Press the new bottle onto the adapter, and swing the bottle into the unit
base.
▶
Putting the cover back on the unit base.
Note
Completely empty packages can be disposed of at recycling facilities.
See also :
Safety data sheet
3.3 Check the OXYGENAL 6 concentration
▶
Test the standard concentration 0.02% H
2
O
2
(200ppm): Fill a measuring
cup (40 ml) with 39 ml water and 1 ml treated water.
43 / 48