2.11 Reprocessing of the service table 1568
Cleaning and disinfection
NOTICE
Non-colour-fast textiles.
Non-colour-fast textiles can cause permanent discoloration.
▶
Remove discolorations with water immediately.
NOTICE
Damage caused by spray disinfection.
Disinfectant gets into cracks and loosens the paint.
▶
Do not use spray disinfection.
▶
Only disinfect by wiping.
Note
The handle on the service table can be taken off to improve the cleaning
properties.
▶
Push the buttons
①
on the bottom of the handle, and pull the handle off in
the direction of the arrow.
1
1
▶
Clean the surfaces with a soft cloth and water, and wipe-disinfect them with
an approved surface disinfectant.
Approved disinfectants:
▪ CaviCide / CaviWipes (Metrex)
▪ Microcide AF Liquid (Schülke & Mayr)
▪ FD 322 (Dürr Dental)
▪ Incidin Liquid (Ecolab)
Sterilisation
Norm Trey and non-slip mat can be sterilised.