Care instructions ESTETICA E70 Vision / E80 Vision
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
12 / 48
2.4.2 Preparing the instrument holder
Note
The handpiece tray with 5-handpieces holder (
Mat. no. 30014236)
and
with 6-handpieces holder (
Mat. no. 30014251
) can be removed for easier
cleaning.
Cleaning and disinfection of the instrument holder
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not treat the instrument holder in a washer disinfector.
▶
Place the handpiece hoses in the designated places of the hygiene centre
during the cleaning and germ reduction of the handpiece holder.
▶
Pull the latches evenly outward and carefully remove the handpiece tray.
▶
Clean the handpiece tray under running water.
▶
Wipe-disinfect the instrument holder.