Care instructions ESTETICA E50 Life
3 Replacing and checking cleaning and germ reduction agents | 3.1 Replenish OXYGENAL 6
48 / 58
3 Replacing and checking cleaning and germ reduction
agents
3.1 Replenish OXYGENAL 6
The only permissible agent for use in KaVo water germ reduction systems is
KaVo OXYGENAL 6 (
Mat. no. 0.489.3451
). The OXYGENAL 6 consumption de-
pends on the water consumption during the treatment of patients and the num-
ber of intensive germ reductions.
When the OXYGENAL 6 container is empty, you hear an acoustic signal every
10 seconds.
Note
Replenish KaVo OXYGENAL 6 only if there is a beep every 10 seconds. The
signal can be turned off as follows: Press the "Intensive germ reduction" but-
ton briefly. The next time the treatment centre is turned on, the beep is is-
sued again.
▶
Remove the tumbler filler
①
and cover
②
. Without patient part assembly kit
with water block DVGW: Take off the lid.
with patient part assembly kit / without patient part assembly kit
▶
Fill-up with KaVo OXYGENAL 6 (max. 1 litre).