Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.6 Preparation of the patient unit
2.6.2 Preparation of the tumbler holder and filler
NOTICE
Water in the device.
Material damage from leaking water upon actuation of the "Tumbler filler" or
"Bowl flush" keys.
▶
Make sure that the unit is turned off before removing the tumbler filler or
the spittoon bowl.
▶
Always check the correct seating of tumbler filler and spittoon bowl before
turning on the unit.
Cleaning and disinfection of the tumbler holder and
filler
Note
When assembling, make sure that the O-ring is present on the tumbler filler.
Note
Inspect the O-ring (
Mat. no. 0.200.6128
) for damage and exchange if ne-
cessary to prevent water or oxidation damage.
▶
Remove the tumbler filler
①
(
Mat. no. 0.764.8541
) by pulling upwards.
▶
Take off the tumbler holder
③
(
Mat. no. 07646153
).
▶
Clean the tumbler holder and tumbler filler under running water.
▶
Wet the O-ring
②
(
Mat. no. 02006128
) with silicone grease (
Mat. no.
10006403
).
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not subject the instrument holder to washer disinfector treatment.
31 / 58