Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
NOTICE
Damage caused by over-heating.
Damage to the gripping sleeve.
▶
Immediately remove the parts from the steriliser after the sterilisation cycle
is completed.
Storage
▶
Reprocessed products should be stored protected from dust with minimum
exposure to germs in a dry, dark and cool place.
2.4.5 Preparation of the instrument hoses
Cleaning and disinfection of instrument hoses
▶
Clean the hoses and couplings with a cloth and water.
NOTICE
Damage caused by spray disinfection.
▶
Do not subject instrument hoses to spray disinfection.
▶
Subject instrument hoses to wipe disinfection.
2.4.6 Servicing the turbine return air filter
Cleaning the turbine return air filter
Note
The liquid collector must be checked weekly.
▶
If liquid is in the filter , drain the liquid and replace the O-rings of the mul-
tiflex couplings.
▶
If necessary, replace the turbine return air filter (
Mat. no. 02003098
) and
clean the filter holder.
17 / 58