Care instructions ESTETICA E50 Life
2 Setup methods according to ISO 17664 | 2.4 Preparation of the dentist unit
12 / 58
▶
Pull the snap-in buttons outward evenly and carefully remove the instru-
ment holder.
▶
Clean the instrument holder under running water.
▶
Wipe-disinfect the instrument holder.
Sterilize the instrument holder
NOTICE
Damage caused by sterilisation.
▶
Do not sterilise the instrument holder.
2.4.3 Reprocessing of the instrument arm S table
Note
The instrument arm S table (
Mat. no. 1.007.4755
) can be removed to facil-
itate cleaning.
Cleaning and disinfection of the instrument arm S table
NOTICE
Damage caused by treatment in a washer disinfector.
▶
Do not sterilise the instrument arm S table in a washer disinfector.
Note
Neither the rinsing program nor the intensive germ reduction can be carried
out when the instrument arm is removed.
▶
Switch off the unit at the main switch.