Instructions for use ELECTROmatic TM and TMM/TMC
5 Startup | 5.4 Making the device settings
▶
Place the airflow measuring tube
①
(
Mat. no. 0.411.4441
) on the motor.
▶
Press the foot control to start the motor.
▶
Adapt the system pressure of the treatment centre appropriately such that
the cooling airflow is 6 to 9 Nl/min (upper edge of sphere
②
).
▶
Comply with the limits for the system pressure from the technical specifica-
tions for media.
See also:
2
3.6 Technical specifications of the ELECTROmatic, Page 27
5.4 Making the device settings
▶
Press the control element, "Settings", to start the "Settings" menu.
▶
Press the arrow control keys to select the various parameters.
▶
Press the Plus control element to increase a value.
or
▶
Press the Minus control element to decrease a value.
ð
The settings are always saved instantaneously.
▶
Press the "Back" control element to exit from the "Settings" menu.
The following device settings can be changed according to need and/or dis-
played when the device is being put into service.
Display information
Setting
The LUX brightness can be adjusted between 0 and
4. "0" means that the LUX light is switched off.
The LUX afterglow time of the straight or contra-
angle handpiece can be adjusted between 0 and 9
seconds. "0" means that the LUX light does not af-
terglow.
The shifting of the files in the file editor can be
switched on or off.
Press the "ON/OFF" control element to switch the
shifting of files in the file editor on or off.
In the default setting, the shifting of files is switched
off (OFF).