Servicing the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a week.
See also: Instructions for Use KaVo QUATTROcare PLUS 2124 A
Note
Handpieces must be taken off the service couplings before the chuck service can
be started and performed.
▶ Remove the service coupling chuck from the side hatch of the QUATTROcare
PLUS and attach it to coupling service point four, on the far right. A MULTIflex
adapter must be mounted there.
▶ Press the handpiece together with the guide bush of the chuck to be serviced
against the tip of the service coupling chuck.
▶ Press the button showing the chuck service symbol.
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start theservice procedure.
Option 2: After three minutes with no service procedure running, the device auto‐
matically switches back to normal service mode.
See also: Care with KaVo QUATTROcare PLUS
7.6
Packaging
Note
The sterilisation bag must be large enough for the handpiece so that the bag is
not stretched.
The quality and use of the sterilisation packaging must satisfy applicable stand‐
ards and be suitable for the sterilisation procedure!
▶ Weld each medical device individually in a sterilised item package!
7.7
Sterilisation
Sterilisation in a steam steriliser (Autoclave) EN 13060/ISO 17665-1
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Before each sterilisation cycle, service the medical device with KaVo care
products.
Instructions for use COMPACTtorque 636 - 0.553.0810 COMPACTtorque 636C - 0.553.0830
7 Reprocessing methods according to ISO 17664 | 7.6 Packaging
19/22