Kavo 3D eXam ® Operators’ Manual
k990400 September 19, 2007
10-14
QA Water Phantom Test
The Water Phantom Test is a noise level test. The HU measurements
are taken at five different regions in the Water volume.
1. Remove Chin Cup and insert Phantom Platform.
2. Fill Phantom with water, approximately one inch from the top.
3. Carefully place Phantom on platform.
4. Using the Alignment Lasers such that the water bath is in the
center of rotation and the horizontal laser line is located across
the center of water depth.
5. Start test by selecting
File > New Patient
from the Main menu.
6. From the Patient Information Panel, click
Add
or right click an
existing image from database and select
Acquire Now
.
If adding a new patient, enter
Water Calibration
in the
First
Name
field and
Noise Test
in the
Patient ID
field.
7. Click
OK
to close the Patient Information box.
8. From the Volume tab in the Acquisition window, select the
following:
Exposure:
Large
Height of Reconstruction Volume:
Landscape 13cm
Resolution:
.4 voxel
9. Click
Preview
to check the phantom position.
Summary of Contents for 3D eXam
Page 30: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 5 8...
Page 46: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 6 16...
Page 90: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 9 12...
Page 126: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 11 10...
Page 161: ...k990400 September 19 2007 B 7...
Page 162: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 8...
Page 163: ...k990400 September 19 2007 B 9...
Page 164: ...Kavo 3D eXam Operators Manual k990400 September 19 2007 B 10...