Instructions for use

Spare cartridge for MASTERtorque -

2.000.2288, 2.000.2266

Always be on the safe side.

Summary of Contents for 2.000.2266

Page 1: ...Instructions for use Spare cartridge for MASTERtorque 2 000 2288 2 000 2266 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...oduct description 3 2 1 Purpose Proper use 3 2 2 Technical data 4 3 Changing the replacement turbine 5 4 First use 7 5 Function test 8 6 Tools and consumables 9 Instructions for use Spare cartridge fo...

Page 4: ...y for damage arising from repairs that have been carried out inappropriately It is permissible to have the replacement turbine exchanged by Technicians at KaVo branches throughout the world Technician...

Page 5: ...lacement part for the indicated turbines and any kind of improper use is impermissible and can be hazardous Proper use According to these regulations this medical device may only be used for the de sc...

Page 6: ...nt turbine assembled 2 x O rings 2 x spring washers Ring mounted Mat no 1 009 5301 Stamp for brake ring Mat no 3 000 0214 1 x wrench Mat no 0 411 3053 Replacement turbine MASTERtorque without wrench M...

Page 7: ...counterclockwise direction Take out defective rotor and brake ring Remove defective rotor and brake ring from the head casing using the short end of the stamp Remove the O rings Remove the O rings fro...

Page 8: ...ngs Replace both O rings Spray O rings with KaVo Spray and insert them in the head housing and cov er Insert the replacement turbine DANGER Not maintaining a torque of 100 Ncm when tightening the cove...

Page 9: ...ear Ensure that the cooling air is dry clean and uncontaminated in accordance with EN ISO 7494 2 Spray the KaVo recommended care product through the medical product and attach the medical product to t...

Page 10: ...medical device Do a test run under axial and radial load with no patient contact Tighten the cover again with a torque of 100 Ncm 6 Ncm Note If the unit runs poorly or heats up in untypical manner pl...

Page 11: ...200 6172 Spring washer 1 003 5900 Stamp for brake ring 3 000 0214 Material summary Mat No Instrument stand 2151 0 411 9501 Insert for medical device 0 411 9902 Cleanpac 0 411 9691 KaVo Torque wrench 1...

Page 12: ......

Page 13: ...1 010 1086 kb 20130114 02 en...