![KaVo 1.009.3630 Instructions For Use Manual Download Page 25](http://html1.mh-extra.com/html/kavo/1-009-3630/1-009-3630_instructions-for-use-manual_658102025.webp)
Instructions for use MASTERmatic LUX M25 L - 1.009.3630, MASTERmatic LUX M05 L - 1.009.3640
7 Reprocessing steps in accordance with ISO 17664 | 7.7 Storage
25 / 28
The medical device has a max. temperature resistance of up to 138
o
C (280.4
o
F).
Sterilisation parameters:
Sterilisation of the product is approved for the following “wrapped” steam steril-
isation cycles:
▪ Steriliser with triple pre-vacuum:
– 3 minutes at 135 °C (275 °F)
Drying time: 16 min.
▪ Steriliser using the gravity method:
– 10 minutes at 135 °C (275 °F)
Drying time: 30 min.
– at least 60 minutes at 121 °C (250 °F)
Drying time: 15 min.
4
Remove the medical device from the steam steriliser immediately after the
sterilisation cycle is completed.
4
Use according to the manufacturer's Instructions for Use.
7.7 Storage
Reprocessed products must be stored appropriately protected from dust in a
dry, dark and cool space with a low germ level.
NOTE
Comply with the expiry date of the sterilised items.