![KaVo 1.003.7708 Instructions For Use Manual Download Page 45](http://html.mh-extra.com/html/kavo/1-003-7708/1-003-7708_instructions-for-use-manual_658086045.webp)
▶
Spray the disinfectant on a cloth, then thoroughly wipe down the med‐
ical device and leave the disinfectant to soak in according to the in‐
structions from the disinfectant manufacturer.
▶
Follow the instructions for use of the disinfectant.
7.3.2 Disinfection: Manual disinfection - internal
The efficacy of manual internal disinfection must be demonstrated by the
manufacturer of the disinfection agent. With KaVo products, use only dis‐
infection agents that have been released by KaVo with respect to the com‐
patibility of materials (e.g. WL-cid / made by ALPRO).
▶
Immediately after internal disinfection, lubricate the KaVo medical de‐
vice immediately with care agents from the KaVo care system.
Setup methods according to DIN EN ISO 17664
43