5 Operation | 5.1 Connecting the motor to the supply hose
14 / 26
5 Operation
Note
At the beginning of each workday, the water-conducting systems should be
rinsed for at least 2 minutes (without transmission handpieces being at-
tached) and if there is a risk of contamination from reflux or back suction, the
system may also need to be rinsed for 20 to 30 seconds after each patient.
5.1 Connecting the motor to the supply hose
▶
Slightly spray the O-rings on the supply hose with KAVO Spray.
▶
Insert the plug-in connector
①
into the motor bearing end plate while not-
ing the position of the plug contacts.
▶
Screw the screw joint
②
tightly in the direction of the arrow.
Note
Details on the hose coupling can be obtained from the manufacturer.
5.2 Regulating the spray water
CAUTION
Hazard from insufficient amount of spray water.
Damage to the cog due to overheating.
▶
Make sure that the motor speed is appropriate for the preparation on hand.
▶
Use the requisite minimal amount of spray water.
▶
Rotate the regulating sleeve
③
right or left to adjust the water portion of
the spray steplessly.
ð
Maximum throughput is achieved when the "KaVo" imprint on the regulating
sleeve is aligned with the button
④
.
5.3 Placing attachment instruments on the motor
All handpieces and contra-angle handpieces can be mounted with the INTRA-
matic connection DIN EN ISO 3964.
CAUTION
Make sure that the handpiece is firmly seated on the motor.
The handpiece detaching inadvertently from the motor during treatment is a
hazard to the patient and user.
▶
Pull on the instrument before each use to check that it is engaged and
locked in place.