Instructions for use
GENTLEsilence 8000BN - 1.000.5700
Always be on the safe side.
Page 1: ...Instructions for use GENTLEsilence 8000BN 1 000 5700 Always be on the safe side ...
Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...
Page 3: ... the O rings on the medical device 12 6 2 Cleaning the spray nozzle 12 6 3 Change the water filter 12 7 Preparation methods according to ISO 17664 14 7 1 Preparation at the site of use 14 7 2 Cleaning 14 7 2 1 Manually cleaning the outside 14 7 2 2 Manually cleaning the interior 14 7 2 3 Mechanically cleaning the exterior and interior 15 7 3 Disinfection 15 7 3 1 Manually disinfecting the exterior...
Page 4: ...d of 12 months from the date of the invoice subject to the following conditions In case of justified complaints KaVo will honour its warranty with a free replace ment or repair Other claims of any nature whatsoever in particular with respect to compensation are excluded In the event of default gross negligence or intent this shall only apply in the absence of mandatory legal regulations to the con...
Page 5: ...orised by KaVo Warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a copy of the invoice or note of delivery The dealer purchase date type and serial number must be clearly evident from this docu ment Instructions for use GENTLEsilence 8000BN 1 000 5700 1 User instructions 1 1 Warranty terms and conditions 3 19 ...
Page 6: ...RNING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury 2 2 Purpose Proper use This medical device is Only intended for dental treatment Any other type of use or alteration to the product is impermissible and can be hazardous The medical device is inten ded for t...
Page 7: ...hould be cleaned serviced and stored in a dry location according to instructions before long periods of nonuse CAUTION Injury or damage due to wear Irregular running noise significant vibration overheating imbalance or insuffi cient grip Stop work and seek service support Note For safety reasons we recommend that the tool holder system be checked an nually after the warranty period expires CAUTION...
Page 8: ...pair and service KaVo products Technicians at KaVo branches throughout the world Technicians specially trained by KaVo To ensure proper function the medical device must be set up according to the re processing methods described in the KaVo Instructions for Use and the care products and care systems described therein must be used KaVo recommends specifying a service interval at the dental office fo...
Page 9: ...d storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold products must be heated to a temperature of 20 C to 25 C 68 F to 77 F Temperature 20 C to 70 C 4 F to 158 F Relative humidity 5 RH to 95 RH absence of condensation Air pressure 700 hPa to 1060 hPa 10 ...
Page 10: ...cm min for spray cooling 4 2 Check the pressure A drive pressure of at least 2 7 to 2 9 bar 39 to 42 psi is required for operating the medical device 2 8 bar 40 psi is ideal The air consumption is approx 47 to 53 Nl min Install the NSK MG 4H test manometer between the hose and coupling Connect test manometer NSK manometer MG 4H to the NSK FlexiQuik cou pling Pressure displayed Induction air 2 8 ba...
Page 11: ...coupling independently snaps into place Pull on it to make sure that the medical device is securely affixed to the cou pling WARNING Release of the medical device during treatment A medical device that is not properly locked can release from the NSK Flexi Quik coupling during treatment Before each use carefully check if the medical device is securely locked on to the NSK FlexiQuik coupling 5 2 Rem...
Page 12: ... cuts Wear gloves or fingerstalls Forcefully press the push button with your thumb and simultaneously insert the cutter or grinder all the way Check that the cutter or grinder is seated by pulling on it CAUTION Hazard from defective chucking system The tool can fall out and cause injury Pull on the tool to check if the chucking system is functioning properly and that the tool is firmly clamped Wea...
Page 13: ...jury Never touch soft tissue with the head or tip since it may be hot and cause a burn After treatment is over remove the cutter or grinder from the contra angle handpiece since injury and infection may result from putting it away when the cutter or grinder are inserted Instructions for use GENTLEsilence 8000BN 1 000 5700 5 Operation 5 4 Removing the milling tool or diamond grinder 11 19 ...
Page 14: ...he spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 0931 if necessary Check the water filter and exchange if necessary Use the nozzle needle Mat no 0 410 0921 to free the water passage in the spray nozzles 6 3 Change the water filter CAUTION Hazard from insufficient amount of spray water Insufficient spray water can cause the medical device to overheat and damag...
Page 15: ... Pull back the insert slightly Place the sleeve on the insert and screw it tight clockwise Instructions for use GENTLEsilence 8000BN 1 000 5700 6 Troubleshooting 6 3 Change the water filter 13 19 ...
Page 16: ...al device The medical device must be dry when transported for reconditioning Do not place it in a solution or similar 7 2 Cleaning CAUTION Malfunctions from cleaning in the ultrasonic unit Defects in the product Only clean manually or in a thermodisinfector 7 2 1 Manually cleaning the outside Accessories required Tap water 30 C 5 C 86 F 10 F Brush e g medium hard toothbrush Brush off under flowing...
Page 17: ...sinfect in a thermodisinfector or manually 7 3 1 Manually disinfecting the exterior KaVo recommends the following products based on material compatibility Mikrozid AF made by Schülke Mayr liquid or cloths FD 322 made by Dürr CaviCide made by Metrex Consumables required Cloths for wiping off the medical device Spray the disinfectant on a cloth then thoroughly wipe down the medical device and leave ...
Page 18: ...h compressed air until water drops are no longer visible Automatic Drying The drying procedure is normally part of the cleaning program of the thermodisin fector Note Follow the instructions for use of the thermodesinfector compressed air quality according to ISO 7494 2 7 5 Care products and systems Servicing CAUTION Sharp cutters or grinders in the medical device Injury hazard from sharp or point...
Page 19: ... of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the cutter or grinder Servicing the product KaVo QUATTROcare plus spray can KaVo recommends cleaning and servicing the chuck system once a week See the KaVo QUATTROcare instructions for use Remove the cutter or grinder place the spray nipple tip in the opening and spray Subsequently treat ...
Page 20: ...280 4 F KaVo recommends for example STERIclave B 2200 2200P by KaVo Citomat K series by Getinge Autoclave with a pre vacuum for least four minutes at 134 C 1 C 273 F 1 8 F Drying time 20 min Autoclave with the gravitation method for at least 10 minutes at 134 1 C 1 C 273 F 1 8 F Drying time 30 min Autoclave with the gravitation method for at least 60 minutes at 121 1 C 250 F 1 8 F Drying time 30 m...
Page 21: ...ey 1 003 6884 Cleanpac 10 units 0 411 9691 STERIclave bags 0 411 9912 O ring 7 4 x 0 8 1 005 8152 O ring 4 4 x 0 85 1 005 8635 NSK spray head for KAVOspray 1 005 8436 KAVOspray 2112 A 0 411 9640 NSK turbine treatment coupling or QUATTROcare 1 000 8786 QUATTROcare plus Spray 2140 P 1 005 4525 Instructions for use GENTLEsilence 8000BN 1 000 5700 8 Accessories 19 19 ...
Page 22: ......
Page 23: ...1 005 8790 kb 20081009 03 en ...