7
Preparation methods according to ISO 17664
WARNING
Hazard from non-sterile products.
Infection hazard for care provider and patient.
▶ Before first use and after each use, prepare and sterilise the medical device
and accessories accordingly.
7.1
Preparation at the site of use
WARNING
Hazard from nonsterile products.
There is a risk of infection from contaminated medical devices.
▶ Take suitable personal protective measures.
▶ Remove all residual cement, composite or blood without delay.
▶ Recondition the medical device as soon as possible after treatment.
▶ Remove the cutter or grinder from the medical device.
▶ The medical device must be dry when transported for reconditioning.
▶ Do not place it in a solution or similar.
7.2
Cleaning
CAUTION
Malfunctions from cleaning in the ultrasonic unit.
Defects in the product.
▶ Only clean manually or in a thermodisinfector.
7.2.1
Manually cleaning the outside
Accessories required:
▪ Tap water 30 °C ± 5 °C (86 °F ± 10 °F)
▪ Brush, e.g. medium-hard toothbrush
▶ Brush off under flowing tap water.
7.2.2
Manually cleaning the interior
To effectively set up, the inside of the machine must be cleaned automatically in a
cleaning and disinfection unit in accordance with ISO 15883-1.
(The interior of this product is not to be cleaned manually).
Instructions for use GENTLEsilence 8000BS - 1.000.5600
7 Preparation methods according to ISO 17664 | 7.1 Preparation at the site of use
13/19