6.2.3 Cleaning: Manual cleaning - internal
Can only be done with KaVo CLEANspray or KaVo DRYspray.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it
on the corresponding care adapter. Press the spray button three times
for 2 seconds each time. Remove the medical device from the spray
attachment and let the cleaner work for one minute.
▶
Afterwards, rinse for 3-5 seconds with KaVo DRYspray.
See also: KaVo CLEANspray / KaVo DRYspray Instructions for Use
Preparation methods according to ISO 17664
42