Care instructions
KaVo SYSTEM 1056 S/T
Always be on the safe side.
Page 1: ...Care instructions KaVo SYSTEM 1056 S T Always be on the safe side ...
Page 2: ...Sales KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Tel 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com ...
Page 3: ...uidelines 2 1 1 1 Abbreviations 2 1 1 2 Symbols 2 1 1 3 Target group 2 2 Preparation methods 3 2 1 Cleaning 3 2 1 1 Manual cleaning 3 2 1 2 Automatic cleaning 6 2 2 Disinfection 7 2 2 1 Manual disinfection 7 2 2 2 Automatic disinfection 7 2 3 Sterilisation 8 2 4 Filling the water bottle 12 1 14 ...
Page 4: ... instructions TA Technician s instructions STK Safety check IEC International Electrotechnical Commission RA Repair instructions EMC Electromagnetic compatibility 1 1 2 Symbols See Safety Hazard Warning Symbol chapter Important information for users and engineers CE Communauté Européenne mark Products bearing this mark conform to the requirements of the pertinent EC directives applicable European ...
Page 5: ... Damage to product as a result of improper disinfection Malfunctions When using disinfectants adhere to the manufacturer s instructions When disinfecting wipe only Do not immerse product or product parts in water or other fluids Note Do not use solvents or aggressive chemicals 2 1 Cleaning 2 1 1 Manual cleaning Switch off the instrument s main switch Dental chair CAUTION Colour transfer from texti...
Page 6: ...asive cleaning fluid Unit body Clean all exterior surfaces with a soft cloth and a mild non abrasive cleaning fluid Aspiration hoses Aspirate 150 ml min 250 ml max of water Push down the dosing device once with the aspiration handpiece attached or by hand Aspirate DEKASEPTOL gel from the dosing device Remove hose and place in holder 4 14 ...
Page 7: ...tion methods 2 1 Cleaning Repeat procedure for the saliva ejector and spray mist aspiration hose DEKASEPTOL gel basic set Mat no 1 000 7204 DEKASEPTOL gel refill canister Mat no 1 000 7205 Spittoon Add one dose of DEKASEPTOL gel to the spittoon 5 14 ...
Page 8: ...t When the requisite action time has passed press the spittoon button The spittoon and drain are thoroughly rinsed Filters Remove aspiration hose from holder Hold a paper towel beneath the filter receptacle on the casing Pull filter out of casing Fit new filter Mat no 0 763 3423 into receptacle 2 1 2 Automatic cleaning Not applicable 6 14 ...
Page 9: ... disinfection Allowed disinfectants S M Microcide Dürr FD 322 Follow the manufacturer s instructions for use 2 2 2 Automatic disinfection Spittoon Rotate spittoon anti clockwise and lift off Place the spittoon in a thermal disinfector and disinfect at 95 C Note Adhere to the manufacturer s instructions 7 14 ...
Page 10: ...Autoclaves with a subsequent vacuum ensure dryness In ad dition drying can be accelerated with a 10 minute drying phase with the autoclave door open CAUTION Damage to product as a result of sterilisation Faulty instrument tubing Do not sterilise instrument tubing Instruments For sterilising instructions see the user instructions supplied with the respective instruments 3 function syringe cannula p...
Page 11: ...TEM 1056 S T 2 Preparation methods 2 3 Sterilisation Detach cannula from 3F handpiece at an angle Unscrew the 3F handpiece cannula in an anti clockwise direction and in a straight line Sterilising the probes in an autoclave 9 14 ...
Page 12: ...an initial va cuum for at least 4 minutes 134 1 Autoclave using the gravitation method at least 10 minutes 134 1 Autoclave using the gravitation method at least 60 minutes 121 1 Once it has been sterilised fit or screw the cannula back on Standard tray Sterilise the standard tray in the autoclave at 135 C for at least 3 minutes 10 14 ...
Page 13: ...te Sterilise the aspiration cannulae in accordance with the instructions supplied To clean the connection pieces hold under running water and clean with a brush Note Connection pieces can be washed in a thermal disinfector at up to 95ºC and ste rilised at up to 135ºC Cannulae When sterilising the cannulae adhere to the manufacturer s instructions supplied 11 14 ...
Page 14: ...ottle can slowly dissipate now CAUTION Improper cleaning of water bottle Destroys the water bottle material When rinsing the water bottle never use water that is above 50 C in tempera ture The water bottle is not dishwasher safe Remove protective hose from the tube of the Oxygenal doser Pass tube through coarse filter opening on the Oxygenal bottle Firmly twist the Oxygenal doser on the Oxygenal b...
Page 15: ...e Rotate opening in lid to the 1 5 litre fill level The lid locks into the fill position Place Oxygenal bottle on a flat surface Push the lid down until the dosage quantity exceeds the fill level The fluid exceeds the fill level the excess is sucked back out to ensure an exact dosage 13 14 ...
Page 16: ...nal bottle in an almost vertical position and fill the water bottle Note If sediment can be seen in the dosing cap this is not contamination it is an in gredient that forms deposits following water condensation if left for a long time These deposits dissolve again after several dosage procedures 14 14 ...