Instructions for use MULTIflex LUX coupling Midwest - 0.553.1390
7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
14 / 20
7 Reprocessing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from contaminated products.
Contaminated products are associated with an infection hazard.
▶
Take suitable personal protective measures.
▶
Reprocess the medical device as soon as possible after treatment.
▶
The medical device must be dry when transported to reprocessing.
▶
To minimise the risk of infection during reprocessing, always wear protect-
ive gloves.
▶
Remove the tool from the medical device.
▶
Remove all residual cement, composite or blood immediately.
▶
Do not place in solutions or similar substances.
7.2 Manual Reprocessing
NOTICE
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶
For manual cleaning only.
7.2.1 Manual cleaning - external
Accessories required:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush, e.g. medium-hard toothbrush
▶
Brush off under flowing tap water.
7.2.2 Manual cleaning of the inside
▶
Rinse the spray air and spray water tube with tap water.
7.2.3 Manual disinfection - external
NOTICE
Malfunctioning from using a disinfectant bath or chloride-containing
disinfectants.
Defects in the device.
▶
Disinfect manually only!