Errors and messages
Instructions for use • TELE PACK + • CQH848_EN_V2.0_08-2022_IFU_CE-MDR
51
10 Errors and messages
10.1 Troubleshooting
Symptom
Possible causes
Actions
Loss of image for longer than
4 seconds
Defibrillator discharged.
Switch the device off and
back on again.
LED temperature above 80°C
Lamp overheated
Allow the lamp to cool
down (to below 75°C)
LED temperature above 90°C
Lamp is switching off after a few
minutes
Lamp overheated
Allow the lamp to cool
down (to below 80°C)
LED temperature above 100°C
Lamp overheated
Turn off the lamp and let it
cool down (to below 75°C)
Device will not start up
Fan malfunction
Contact Service
Device temperature above 80°C
The device switches itself off at
temperatures above 90°C
Device overheated
Allow the device to cool
down (to below 75°C)
Contact Service
Update failed
Restart the device
Repeat the procedure
Contact Service
Incompatible camera head
Replace the camera head