Ver. 2
20160719
Handheld Odor Meter
OMX-ADM
Operation Manual
www.
GlobalTestSupply
.com
Find Quality Products Online at:
[email protected]
Page 1: ...Ver 2 20160719 Handheld Odor Meter OMX ADM Operation Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 2: ...g Mode 12 3 1 1 Measurement Preparation 12 3 1 2 Measurement 13 3 2 Memory Sampling Mode 13 3 2 1 Measurement Preparation 13 3 2 2 Measurement 13 3 2 3 Memory Overflow 14 3 3 Peak hold Function 15 3 4 Zero point Function 15 4 Data Filing Software 16 4 1 How to Download Software 16 4 2 How to Install Software 16 How to Install USD Drivers Software 17 4 3 Preparation 18 www GlobalTestSupply com Find...
Page 3: ... 1 Replacement of AA Alkaline Batteries 20 6 2 Replacement of Button Battery 20 6 3 Maintenance of Air Purifying Unit 21 6 4 Replacement of Filter for Suction Nozzle 21 6 5 Troubleshooting 22 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 4: ...tents and find you received the items as below in good condition The instrument kit contains Name Q ty OMX ADM Base Unit 1 Suction Nozzle 1 AC Adaptor 1 AA Alkaline Battery 4 USB Connection Cable 1 Air Purifying Unit 1 Activated Carbon 50cc bag 1 Replacement Filter for Air Purifying Unit 4 Replacement Filter for Suction Nozzle 5 Operation Manual 1 www GlobalTestSupply com Find Quality Products Onl...
Page 5: ... adaptor Do not use in explosive atmosphere Do not use the device where combustible and or explosive gasses exist The device is not designed for explosion proof Do not dismantle the base unit www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 6: ...s directly The unit is exposed to direct sunlight Heating occurs due to heat radiation Do not use or store the unit at a place where the following gases exist Extremely concentrated odor Cigarette smoke and or oil mist Silicon compound gas included in hair spray adhesive and silicon rubber etc High concentrated chloride acidic substances corrosive gas and chlorofluorocarbon etc Exchange new batter...
Page 7: ...7 1 4 Unit Dimension Unit mm www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 8: ...measured continuously Odor Intensity Odor Strength is converted into ammonia odor intensity by the scale of 2 5 to 5 0 in 0 1 unit based on odor concentration of measurement environment NOTE Odor Intensity is determined based on Japanese Regulation that allows evaluating odor level by 6 scales Peak hold Maximum odor strength level can be retained Memory Sampling Measurement Mode Up to 511 measurin...
Page 9: ... approx Moderate Measureable for 1 to 2 hours approx Low Change battery immediately No battery USB Shows USB Cable is connected Odor Intensity Displays odor intensity of ammonia gas concentration based by the scale of 2 5 to 5 0 in 0 1 unit Hold ENTER for approx 3 seconds to turn ON or OFF Data No Shows number of saved data Memory Displays remaining memory capacity from 0 to 100 Sampling Rate Sele...
Page 10: ...or if necessary Suction Nozzle Takes sampling air and or object gases Exhaust Nozzle Exhaust sampling air and or object gases USB Port Connect attached USB cable for PC connection Battery Box Open Close to replace batteries Hook Pin Use for straps Calendar Settings 1 Hold and turn on the device at the same time to set calendar 2 When Year is blinking change to the arbitrary number 3 Press to confi...
Page 11: ...arbon when deodorization using air purifying unit is not effective Replacement Filter for Air Purifying Unit Replace filter when deodorization using air purifying unit is not effective Replacement Filter for Suction Nozzle Replace filter when gets dirt USB Connection Cable Use to transfer measuring data from the unit to PC www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSu...
Page 12: ...ptor or 4 Turn on the unit at clean ambient It takes about 10 min for the unit initializing and start measuring NOTE The unit sets zero point when the initializing is done NOTE To get the zero point stable initializing at clean environment is recommended NOTE To get Zero point forcibly use air purifying unit See 3 4 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply co...
Page 13: ...mory Sampling Mode 3 2 1 Measurement Preparation Refer to Page 11 3 2 2 Measurement 1 Turn on the unit and wait for about 10 minutes until Ready stop blinking and the initializing is done 2 Press until sampling rate on LCD display starts blinking 3 Select sampling rate by pressing 4 Press to finish selecting sampling rate 5 Press one time and twice to save data www GlobalTestSupply com Find Qualit...
Page 14: ...tarts blinking and data number becomes 0 To delete saved memory press and at the same time while the unit is turned off Keep pressing and and turn on the device NOTE Saved memory in the unit will not be deleted automatically when it is transferred to PC www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 15: ...initialization is done After 10 minutes initializing the unit will display Odor Strength 0 baseline automatically regardless of the measurement environment If you need to set the baseline accurately connect Air Purifying Unit to purify the air forcibly 1 Press until calendar on LCD display 0 and starts blinking 2 Press to proceed or press to cancel NOTE If you set zero point forcibly under dirty e...
Page 16: ...PDrivers aspx 2 Install the software NOTE If PC CPU is 32bit select CP210xVCPInstaler_x86 exe If CPU is 64bit select CP210xVCPInstaler_x64 exe NOTE Warning message appears if incorrect file is selected 3 Confirm Device Manager displays Silicon Labs CP210x USB to UART Bridge COM after the installation 4 Confirm assigned COM number www GlobalTestSupply com Find Quality Products Online at sales Globa...
Page 17: ... Open Software 1 Double click shortcut icon on desktop or open the file from program files 2 Confirm main screen appears 3 Read Handheld Odor Meter Data Filing Software Manual for details www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Page 18: ...to 999 Odor Intensity 2 5 to 5 0 Response Time Ammonia 2ppm Reaches 90 responsiveness within 20 seconds Ammonia 10ppm Reaches 90 responsiveness within 20 seconds Power Source 4 AA alkaline batteries or AC adaptor Consumed Power 1 100mW approx with AA alkaline batteries Storage Temp Range 0 to 40 degree in C No dew condensation Dimensions 74 W x 167 50 L x 35 D Memory 32 732 data or 511 files www G...
Page 19: ...s approx after purchasing Confirms when calendar or time on LCD display is incorrect 1 Turn off the unit and remove AA alkaline batteries NOTE Do not connect AC adaptor 2 Push the round part of button battery firmly until the left part of button battery moves up 3 Take out the battery NOTE Do not use metal tools or easily broken stools like screwdrivers or pencils 4 Insert the new battery www Glob...
Page 20: ...3 Replace bottom filter and fill in activated carbon up to the line NOTE Do not exceed the line when replace activated carbon 4 Replace top filter and top wire gauze 6 4 Replacement of Filter for Suction Nozzle Twist Suction Nozzle and replace filter if necessary Top Top Wire Gauze Top Filter Arrow Label Activated Carbon Bottom Filter Bottom Wire Gauze Bottom Filter Suction Nozzle Do not exceed th...
Page 21: ...nking Odor concentration is too high and overflowed Try to lower the gas concentration Time on LCD display keeps blinking Memory is overflowed Delete data Sampling rate cannot be selected Memory is overflowed Delete data Data becomes 0 Memory is overflowed Delete data SErr is shown on LCD display Sensor malfunction Contact us if the message keeps being displayed PErr is shown on LCD display Built ...