Kangaroo® 224 Feeding Pump
NORMAL OPERATION
Plug the pump into AC wall outlet.
BATTERY OPERATION
Unplugging the pump from the AC line will automatically put the pump on battery
powere. to recharge the battery, plug the pump into an AC wall outlet. Pump will
automatically begin charging.
These general instructions are for inserting the Kendall Healthcare Pump
Administration Sets into the Kangaroo 224 Feeding Pump.
1.
Fill the feeding bag prior to setting up pump.
2.
Close roller clamp on pump set.
3.
Press “ON”
4.
Wait while pump performs systems check.
5.
When display reads 0, set rate by turning knob on side of pump.
6.
Insert drip chamber into drip chamber guides.
7.
Open roller clamp, fill entire line with fluid, close roller clamp.
8.
Attach distal connector to feeding tube.
9.
Grasp the silicone portion of tubing and gently stretch around the pumps rotor.
Insert the pump set retainer disc into retainer lock on the pump. Thread tubing
through tubing guide.
10.
Open roller clamp, press “START/HOLD”. While pump is running, dots will
flash across display.
TO CHANGE RATE:
1.
Press “START/HOLD”.
2.
Set new rate by rotating knob on side of pump.
3.
Press “START/HOLD”.
Directions For Use