Before use

3

Check

Particle filter is not dirty;  Water trap & flue probe empty of water;  

Water trap & red plug fitted to analyser;  Flue probe hose connected to flue gas inlet;  Flue 

probe temperature plug connected to T1 temperature connection

 

Fresh air purge

Position flue probe in fresh air & press “On/Off”  

Analyser auto-calibrates within 30 seconds

Select “Ratio” on dial - in fresh air CO reading = 0ppm

Select “O

2

/Eff” on dial - in fresh air O

2

 reading = 20.9%

Select “Status” on dial to view

 

To charge NiMH rechargeable batteries

Switch off

Connect charger - charging indicator illuminates

Connections

2

Battery charger

Watertrap drain 

with red plug

Flue probe 

temperature T1

Inlet temperature T2

Flue gas inlet

Red connector 

- plugs into flue 

gas inlet

Flue probe temperature plug

- plugs into T1

Narrower (+ve) pin must be on 

right hand side

Flue probe hose