TECHNICAL DESCRIPTION
ULTRAFLOW® 54 DN150-300
Kamstrup A/S
·
Technical description
·
5512-876_M1_GB_12.2020
38
8.4
Pulse emission
Pulses are emitted at 1-second intervals. The number of pulses to be emitted is calculated every second. Pulses are
emitted in bursts with a pulse duration of 2
…
5 ms and pauses depending on the current pulse frequency. The duration
of the pauses between the individual bursts is approx. 30 ms.
The transmitted pulse signal is the average determination of a series of flow measurements. This means that there will
be a transient phenomenon until correct flow signal has been obtained during start-up. Furthermore, this brings about a
pulse tail of up to 8 seconds in case of sudden hold.
8.5
Supply and power consumption of ULTRAFLOW®
Normally, ULTRAFLOW® is supplied by the connected MULTICAL® calculator. When ULTRAFLOW® is supplied in any other
way, for example via direct connection in a flow bench, the following applies:
Supply voltage of ULTRAFLOW®:
3.6 VDC ±0.1 VDC
The current consumption of ULTRAFLOW®:
Max average
50 µA
Max current
7 mA (max 40 ms)