Easidew PRO XP User’s Manual
Kahn Instruments
11
INSTALLATION
2.5
Preparation of the Sensor Cable
The sensor cable is not supplied with the Easidew PRO XP, but
there are quantity 2 ferrules supplied. Maximum conductor size
is 0.75mm².
Correct preparation of the cable conductors is essential to insure a reliable connection
to the sensor terminals.
1.
Strip the insulation from each conductor to 8mm, without damaging any
wire strands.
2.
Using a crimping tool to correctly attach each ferrule.
Figure 13
Wire and Crimp Details
2.5.1 Terminal Block Connection - Easidew PRO XP EX1 (Non-Display)
1.
Remove the enclosure lid by carefully unscrewing counter-clockwise
2.
Remove the 4-way terminal block from its header on the pcb, being careful
not to put any strain onto the ground wire already attached.
3.
Mount each of the power and return wires into positions 2 and 4 as
shown, and tighten with a flat-blade screwdriver (min torque
0.2 lb-ft
(0.25Nm)).
Figure 14
Terminal Block Connection - Easidew PRO XP EX1 (Non-Display)