Figure 38 Results of verification sweep (Return Loss vs Distance screen)
If the Return Loss vs Distance plot does not satisfy the above criteria, it is possible that your RTF module requires
recalibrating. Contact your local Kaelus representative (see Section 9.8) to arrange this.
8.6
Changing the RTF Sweep Settings (iQA)
The default settings in the RTF Analysis window have been chosen to provide a favorable tradeoff between resolution,
sensitivity and measurement speed. However, there may be situations in which the user wishes to change some or all
of these settings. The following sections explain how this is done.
8.6.1
Enabling Configuration Mode
The setup of the instrument can be modified via password access to the configuration mode. To access the
configuration mode:
1.
Press the Config Menu button. A keyboard will be displayed.
2.
Type in the password and press Enter. The screen will now switch to Configuration Mode, indicated by the Config
Unlocked indicator at the top of the screen, as shown in Figure 39.
3.
To return to User mode press the Return button, or the Config Mode indicator.