Criteria
Specification
Overall dimensions, (L x W x H)
860mm x 550mm x 180mm
Weight
53kg maximum
Temperature range
35 – 42°C
User interface
Touchscreen
User interface functions
Digital temperature readout, data logger, temperature
setpoint, calibration, warning for next service
Connections
Mains, CO
2
gas, N
2
gas, Sampler port, Ethernet, Alarm
Alarms
Visual and audible alarm for out of range temperature and gas
Filter (HEPA/ VOC)
Origio Gas Line Filter
IP class
IP30
Overvoltage category
Transient overvoltage II
Pollution degree rating for electrical
equipment
2
Power specifications 100 – 240 VAC
Max consumption
270W
Voltage
L/N/PE AC, 100 - 240VAC Class 1 type B
Frequency
50/60 Hz
Current
3A
Mains supply voltage fluctuations
Up to +/-10 % of the nominal voltage
Fuses 100 - 250V UL Listed
Mains connection
T4.0AL
Ambient conditions
Working temperature and humidity 20 – 30°C. Less than 75% RH (non-condensing)
Transport and storage temperature
and humidity
-5 – 50°C. Less than 95% RH (non-condensing)
Maximum one week at 50°C