Dewert User Manual 2690XLE/2695XLE/2890XL/2895XL – Updated 11.2021/3 MDR Page 6 of 15
Middle part adjustment with models 2890XLE, 2890XL/H, 2895XLE, 2895XL/H
The foot part is adjusted using a gas spring. To operate, press the release lever underneath the middle part in
the direction of the upholstered surface. The middle part is lifted slowly to the positive end position. To lower,
press down the upholstery and operate the release lever at the same time. Once the desired position has
been reached, let go of the release lever. The strength of the gas spring ensures ease of adjustment despite
the patient’s weight. Consequently, when unloaded, more force is required to adjust the middle part
horizontally again.
2.5 Foot part adjustment
Caution: Weight load at the outer foot end: max. 90 kg
Foot part adjustment with models 2690XLE, 2695XLE, 2890XLE, 2890XL/H
The foot part is adjusted using a gas spring. To operate, press the release lever at the end of the foot part in
the direction of the upholstered surface.
The foot part is lifted slowly to the positive end position. To lower, press down the foot part upholstery and
operate the release lever at the same time. Once the desired position has been reached, let go of the release
lever. The strength of the gas spring ensures ease of adjustment despite the patie
nt’s weight. Consequently,
when unloaded, more force is required to adjust the foot part horizontally again.
Lower foot part adjustment with models 2895XLE, 2895XL/H
The lower foot part is adjusted using two gas springs. To operate, press the release lever at the foot part end
underneath the upholstery in the direction of the upholstered surface. The foot part is lifted slowly to the
positive end position. To lower, press down the foot part and operate the release lever at the same time.
Once the desired position has been reached, let go of the release lever.
2.6 Movability (model dependent or optional)
Individually lockable twin castors
The castors can be locked using the foot-operated brake on the castor housings. In this case, the castors can
neither be moved nor rotated. To release, operate the brake on each castor again.
Please note: Standard castors are not electrically conductive = optional equipment.
You can recognise conductive castors via the respective marking = yellow dot on the side of the running
surface or a yellow ring.
With this movability option, access to the patient is improved during examination and treatment thanks to easy
positioning within the room. Transporting patients is not considered as intended use.
Central movability
By operating a lever (on the outside of the table feet), all four castors are activated simultaneously. The
following moving options exist:
Stage 1: The castors of the table are locked and can neither be moved nor rotated.
Stage 2: = centre position: The castors are released and can be moved and rotated, the table can be moved
in all directions.
Stage 3: Three castors are released (= can be moved and rotated). The fourth castor is locked and cannot be
rotated (directionally locked castor), i.e. the wheel rolls in a fixed position and helps to steer the table in the
intended direction.
Please note:
Rotational movement is only prevented when the castor is swivelled parallel to the lying surface.
This then allows the table to be moved in a straight line without pulling off to the side.
Wheel-lifting/fixing mechanism
The wheel-lifting/fixing mechanism allows a combination between fixed and movable table. There are two twin
castors on each short side of the underframe and foot levers at each corner. The foot levers consist of two
ergonomically arranged counter holders.
This way, the table can be lifted and lowered quietly with your foot. The wheel-lifting/fixing mechanism is not
controlled centrally; a foot lever must be operated on each short side of the table in order to lower the table
onto its feet or to place it on the castors. This option ensures the device can be simply relocated; however,
the device is not intended for transporting patients (ground clearance (distance foot - ground) when moving
the table is approx. 14 mm).