K.H. DEWERT 2690XLE User Manual Download Page 11

Dewert User Manual 2690XLE/2695XLE/2890XL/2895XL – Updated 11.2021/3 MDR                              Page 11 of 15 

  

Risk of electrocution! 

Work on the electrical system may only be carried out by qualified and authorised personnel in 
compliance with all relevant provisions and safety regulations! 

 

Foot and hand switches for adjusting the electric motor as well as gas springs are wearing parts whose 
function may be impaired over the years depending on the frequency of use. 
Both can be replaced without too much effort. Please request the corresponding installation diagram, if 
required. Replacement parts can be obtained directly from Dewert. 
 
 

7. Safety devices 

Tables with electromotive adjustment must have an automatic device to deactivate the control elements when 
moving the table. Reactivation of the control elements must be designed in such a way that they cannot be 
triggered accidentally by patients, users or third parties. 
 
The actuator of this table is equipped with an integrated safety device to protect against 
unauthorised/unintentional operation. This device goes into “sleep mode” 3 seconds after the last operation 
and can only be reactivated with a defined switching sequence, the so-called double tap. 
To “wake up” the actuator or the control unit, first press the desired direction of travel for approx. 1 second on 
the control element. After a short waiting time (1-2 seconds), press the desired direction of travel again and 
the actuator can be moved for a maximum of 30 seconds in this direction of travel. 
If the switching cycle of the double-tap function is not observed, the actuator cannot be operated. After 30 
seconds of operation in one direction of travel, the actuator switches off and goes into sleep mode. The 
actuator can still be operated for up to 3 seconds after the last operation to ensure fine adjustment. Within this 
time window, it is possible to move in each direction of travel again for a maximum of 30 seconds. The 
actuator always goes into “sleep mode” automatically 3 seconds after the last operation.  
 
The motor is equipped with a 

safety freewheeling clutch

 as standard: 

When lowering, the motor declutches automatically if an obstacle is encountered, and automatic declutching 
is interrupted. In other words, the active tractive force of the motor no longer acts; instead, simply the weight 
of the upper part of the table is applied. In the event of unforeseen entrapment, the risk of injury is significantly 
limited.  
 

Foot switch cover 

The top of the foot switch is fitted with a protective cover. This makes unintentional or unauthorised operation 
more difficult. 

 
 

8. Reporting obligation 

All serious incidents occurring in connection with the product shall be reported to the manufacturer (

K.H. 

Dewert GmbH

) and to the 

competent authority

 of the Member State in which the user and/or the patient 

resides. 

Member State 

Competent authority 

Web 

Estonia 

Health Board 
Terviseamet 

https://www.terviseamet.ee/en 

Finland 

Fimea 

https://www.fimea.fi/web/en 

Iceland  

Iceland Medicines 
Agency 

https://www.ima.is/ 

Luxembourg 

CNS 

https://cns.public.lu/en 

Malta 

MCCAA 

https://mccaa.org.mt/ 

Switzerland 

Swiss medic 

https://www.swissmedic.ch/swissmedic/de/home.html 

A serious incident is an incident that directly or indirectly had, could have had, or may have had any of the 
following consequences:  

Summary of Contents for 2690XLE

Page 1: ...setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 5 2 1 Table design 5 2 2 Height adjustment 5 2 3 Head part adjustment 5 2 4 Middle part adjustment 5 2 5 Foot part adjustment 6 2 6 Movability model dependent or optional 6 2 7 Additional equipment 7 2 8 Special features of models 2690XLE 26...

Page 2: ...sing the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficiently familiar with its adjustment mechanism or have read and understood the operating manual fully The manufacturer cannot be held responsible for damage caused by or involving unauthorised persons No t...

Page 3: ...e is ready for use upon delivery Remove the power cable from the film packaging on the underframe of the table and connect it properly to a permanently installed mains socket When routing the power cable make sure the cable cannot be crushed rolled over or otherwise damaged Lock the castors optional or the wheel system optional As the operator carry out a thorough and precise function check once t...

Page 4: ...and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white aluminium powder coated RAL 9006 04 grey aluminium powder coated RAL 9007 and the type of selected height adjustment E electromotive H hydraulic The type plate is attached to the underframe on one long side of the table It provides information about key table data The follo...

Page 5: ... time of at least 400 s must be observed If the maximum duty cycle is exceeded an internal thermal switch protective temperature limiter in the motor interrupts the power supply to the actuator After the electric motor has cooled down the thermal switch automatically reconnects the power supply to the actuator Height adjustment with hydraulic height adjustment At tables equipped with hydraulic hei...

Page 6: ...stors The castors can be locked using the foot operated brake on the castor housings In this case the castors can neither be moved nor rotated To release operate the brake on each castor again Please note Standard castors are not electrically conductive optional equipment You can recognise conductive castors via the respective marking yellow dot on the side of the running surface or a yellow ring ...

Page 7: ... the head part from the side The armrests must not be used for sitting They only serve to support the patient s arms 2 8 Special features of models 2690XLE 2695XLE The tables for echocardiography treatment have a cut out on the left hand side in the upholstered surface It can be closed with a cushion The cushion can be pushed out from the bottom upwards a slight amount of pressure is required To c...

Page 8: ...eight may vary depending on the equipment The motor is equipped with a thermal protective switch as standard This causes the motor to cut out if the weight load is too extreme or the motor s duty cycle DC 25 s 400 s 25 s 400 s is exceeded The table should be ready for use again after a 15 minute rest period There is therefore no risk of overloading the motor The mechanics of the tables have been d...

Page 9: ...the separate enclosed sheet The table is equipped as standard with the Dewert hygiene standards which enable optimal cleaning and disinfection Hinge lids are made of the identical upholstery covering All undersides of the upholstery parts are coated with an upholstery covering and can thus be cleaned and disinfected Vent holes underneath the upholstery For the homogeneous foam to recover quickly a...

Page 10: ... proper fit of the screw connections Lubricate the swivel joints and operating levers slightly with low viscosity spray oil if necessary Lubricate the roller guides with a little bearing grease or similar using a brush if necessary Insufficient lubrication is noticeable through noise development A checklist for maintenance technical inspection can be found in the appendix In addition to maintenanc...

Page 11: ...p function is not observed the actuator cannot be operated After 30 seconds of operation in one direction of travel the actuator switches off and goes into sleep mode The actuator can still be operated for up to 3 seconds after the last operation to ensure fine adjustment Within this time window it is possible to move in each direction of travel again for a maximum of 30 seconds The actuator alway...

Page 12: ...in the Dual System in accordance with the German Packaging Act VerpackG and therefore bear the disposal costs meaning you can dispose of the packaging free of charge Product Safety notices o Pay attention to sharp edges and pointed objects o When transporting the table only take hold of the underframe NOT the upper frame o In order to prevent accidents later on the no longer used product must be r...

Page 13: ... purpose of curative and disease treatment examination massage and health therapy Table operation and patient positioning on the table may only be performed by professionally trained persons who have been instructed in its use or who through experience with other similar medical devices have knowledge of its proper use taking into account possible hazards The device must only be moved within the r...

Page 14: ...d supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off to ensure that the table operates smoothly without any collisions or blockages no cable connection can be overstretched crushed or otherwise damaged the motors run without noticeable noise development the end position switc...

Page 15: ...le All necessary toggle screws present Possible movability Castors undamaged freewheeling given Connection to frame undamaged Re tighten all screw connections with central locking system also the grub screws of switch levers Safe braking effect Check with locked brakes by pulling and pushing the table Central Movability Levers must be in contact with the frame tighten the fastening screws Shift le...

Reviews: