Dewert User Manual Heavy-Weight Tables – Updated 11.2021/3 MDR                                                         Page 9 of 14 

Important: 

At tables with hydraulic height adjustment as well as for the gas springs, wipe the piston rod with a soft cloth 
at regular intervals. This prevents dust from entering through the dust lip and preserves the service life of the 
unit. 
 

 

   Caution: 

When cleaning, secure the table against unintentional lowering of the lying surface. 
To do so, place the head part horizontally and disconnect the power plug from the mains socket.  

 
The power plug must not come into contact with water or cleaning agents. 
 
The electrical components must not display any external damage through which liquid could enter. Do not 
clean the table with water jets, a high-pressure cleaner or with a so-called bed washer. Only use moist cloths. 
 
 

6. Maintenance and technical inspection 

The device has been designed and manufactured in such a way that it will operate safely over a very long 
period of time if used as intended and in a correct manner. Depending on the conditions of use, place of use 
and care, the expected operating service life is 10 years or 100,000 drive cycles (double stroke = 1x up and 
down). 
 

Regular maintenance procedures

 are required to ensure patient, user and product safety. They must be 

carried out every two years at the latest. 

 

 

The maintenance procedures can be carried out by qualified personnel/instructed personnel. 
The scope of maintenance includes, for example:  

 

Thorough visual inspection of all components, especially motor and switch with mains supply line or 
hydraulic system 

 

Check all swivel joints for completeness. 

 

Check proper fit of the screw connections.  

 

Lubricate the swivel joints and operating levers slightly with low viscosity spray oil, if necessary. 

 

Lubricate the roller guides with a little bearing grease or similar using a brush, if necessary. 
 

Insufficient lubrication is noticeable through noise development.  
 
A checklist for maintenance/technical inspection can be found in the appendix. 
 

In addition to maintenance in accordance with the legal requirements of DGUV Regulation 3 / IEC 
62353, a technical inspection must be carried out every two years at the latest on tables with 
electromotive adjustment. 
This technical inspection may only be carried out by authorised and trained specialists. 
A checklist for maintenance/technical inspection can be found in the appendix. 

 
Despite regular maintenance/technical inspection, the 

user

 is also responsible for patient safety and proper 

and reliable functioning. As the user, make sure that the table is in good working order prior to each use 
(visual inspection). In the event of any abnormalities, take the table out of service immediately and inform the 
operator. 
 

 

Replace damaged or worn components immediately and do not use the table until it has been repaired. 
 

 

The table complies with the safety regulations prescribed at the time it was placed on the market. Improper 
repairs and structural modifications (disassembly of original parts, installation of third-party parts, etc.) may 
result in hazards for patients and users. In the event of uncoordinated modifications to the table, the 
Declaration of Conformity loses its validity and the warranty shall become null and void. 
We cannot be held liable for damages resulting from uncoordinated modifications. 
Only original Dewert spare parts may be used. 
 

Summary of Contents for 208

Page 1: ...mbols 2 1 2 Applied standards 2 1 3 Safety regulations 2 1 4 Intended purpose 2 1 5 Information on setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 5 2 1 Table design 5 2 2 Height adjustment 5 2 3 Head part adjustment 5 2 4 Movability model dependent or optional 5 2 5 Additional equipment ...

Page 2: ...the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authorised instructed and competent persons who are sufficiently familiar with its adjustment mechanism or have read and understood the operating manual fully The manufacturer cannot be held responsible for damage caused by or involving unauthorised persons No third ...

Page 3: ...e do not lift it at the head part as this may damage the head part and or the release mechanism Hold and lift the table on the left and right hand side of the frame The table must stand securely on its feet or castors on a level flat and solid surface Before use activate the brakes on the castors and make sure they are working properly When transporting the table take hold of the underframe NOT th...

Page 4: ...during maintenance work To disconnect be sure to grasp the plug not the power cable This device must not be modified without the express permission of the manufacturer When transporting the table only take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the choice of frame colour 00 white powder coated RAL 9010 03 white alu...

Page 5: ...ke place via the foot switch or the hand switch Please refer to section 7 The table is lifted or lowered Note on operation The electric motor is to be operated in intermittent duty mode This means that a maximum duty cycle of 25 s must not be exceeded Before switching the motor back on an interruption time of at least 400 s must be observed If the maximum duty cycle is exceeded an internal thermal...

Page 6: ... second hole Then relieve the spring tension on the sleeve To release the paper roll holder proceed in the same way Push handles optional Depending on the actual equipment push handles are located either at the end of the head part or the foot part or on both sides When moving the table always use both hands on the push handles to push pull the table Side guard optional Foldable side guard Folding...

Page 7: ...impact protection Additional foot switch Additional hand switch Foot switch fixation on the underframe 4 Technical data 4 1 Individual models 208 209 7209E 7255E Models Max length mm 1950 1950 1950 1990 Width mm 650 800 800 800 Head part length mm 540 540 500 540 Weight approx depending on equipment kg 60 70 110 130 Min max height mm 540 890 590 930 Fixed height mm 650 650 Adjustment time height m...

Page 8: ...n the meaning of EN 50082 Part 1 and 2 5 Cleaning instructions Upholstery covering There are two different collections of upholstery covering to choose from Skai Pandoria Plus manufactured by Hornschuch Continental in Germany Skai Toronto EN manufactured by Hornschuch Continental in Germany For cleaning and disinfection a selection of agents from various manufacturers has been tested for compatibi...

Page 9: ...wivel joints for completeness Check proper fit of the screw connections Lubricate the swivel joints and operating levers slightly with low viscosity spray oil if necessary Lubricate the roller guides with a little bearing grease or similar using a brush if necessary Insufficient lubrication is noticeable through noise development A checklist for maintenance technical inspection can be found in the...

Page 10: ...the control unit first press the desired direction of travel for approx 1 second on the control element After a short waiting time 1 2 seconds press the desired direction of travel again and the actuator can be moved for a maximum of 30 seconds in this direction of travel If the switching cycle of the double tap function is not observed the actuator cannot be operated After 30 seconds of operation...

Page 11: ...sal Packaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Please observe the local regulations for waste disposal and preferably recycle the materials As a manufacturer we are licensed as a participant in the Dual System in accordance with the German Packagin...

Page 12: ...s use or who through experience with other similar medical devices have knowledge of its proper use taking into account possible hazards Equipped with the options of movability not wheel lifting fixing mechanism side guards and push handle the intended purpose of the table is extended and also provides for the ideal positioning of patients for the purpose of transport to pre treatment or post trea...

Page 13: ...d supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off to ensure that the table operates smoothly without any collisions or blockages no cable connection can be overstretched crushed or otherwise damaged the motors run without noticeable noise development the end position switc...

Page 14: ...le All necessary toggle screws present Possible movability Castors undamaged freewheeling given Connection to frame undamaged Re tighten all screw connections with central locking system also the grub screws of switch levers Safe braking effect Check with locked brakes by pulling and pushing the table Central Movability Levers must be in contact with the frame tighten the fastening screws Shift le...

Reviews: