4VetJ Ultrasound Scanner User’s Manual
Penetration Depth
The depth of penetration is set with the help of the [Depth
–] and [Depth +] buttons. The depth of scanning is set
depending on the location of the examined organs. The
depth of the beam penetration depends in its frequency.
The higher the frequency is, the lower its range of
penetration is.
You can also change the penetration depth by moving a
finger or the stylus in the active field at the right edge of the
sector of scanning.
Focusing the Beam
The [Focus -] and [Focus +] buttons are used to set the
focus of the ultrasound beam. The image presented on the
screen is the sharpest on the level of the beam focusing.
The level of focus is indicated by a red triangle located on
the left side of the sector. To obtain the best effects, set
the focus on the level corresponding to the location of the
examined object.
Focus adjustment can also be performed using a finger or
the stylus in the vertical active field located on the left side
of the sector.
The scanner also has automatic dynamic focusing, so the
image is sharp in the whole range of the scanning.