8 Startup and parameterization
66
8.7.1
Adjusting the current output AO1
This function allows for the adjustment of the signal values emitted at the analog output.
1. If you do not wish to change any further parameters, go to the END function in the test menu and
press the
key in order to save the settings and return to the process level.
CAL. AO1
OFFSET
OF = 4.05
The transmitter generates a current of 4 mA.
1. Measure the current emitted at the 4 to 20 mA output using a multi-
meter.
2. Enter the value displayed by the multimeter. The permitted offset
range is 3 to 5 mA.
3. Confirm
SPAN
SP = 19.95
The transmitter generates a current of 20 mA.
1. Measure the current emitted at the 4 to 20 mA output using a multi-
meter.
2. Enter the value displayed by the multimeter. The permitted span
range is 18.50 to 21.50 mA.
3. Confirm
VALID N/Y
1. Confirm the entered value or not.
CAL. AO1