8 Startup and parameterization
40
8.6.3
Entering the K-factor for the fitting (406090) used
The device calculates the medium's flow rate in the pipeline with the K-factor for the fitting (406090)
used.
You can enter the K-factor for the fitting used here. The device can also determine the K-factor in a cal-
ibration procedure (Teach-In),
s see chapter 8.6.4 "Determining the fitting's K-factor with a calibration
procedure (Teach-In)", page 41.
1. If you do not wish to change any further parameters, go to the END function in the parameters menu
and press the
key in order to save the settings and return to the process level.
NOTE!
The device starts using the new K-factor as soon as the function SAVE YES is confirmed when leaving
the parameters menu.
K-FACTOR
K = 10.000
K = 2.8500
The display shows the
K-factor for the
fitting (406090) used.
1. Confirm the parameter.
1. Adjust the K-factor for the fitting
(406090) used (value between
0.0001 and 9999.9). Changing
a numerical value,
see
chapter 8.4 "Using the naviga-
tion keys", page 35.
2. Confirm the displayed value.
OUTPUT
RETURN
VALID N/Y
Displayed only if the K-factor has
been changed.
1. Confirm the entered K-factor or
not.