
11 Calibrating a sensor with a standard signal
86
11.6 Chlorine measurement operating mode, pH-
compensated
11.6.1 Final value calibration
The transmitter is in "Measuring mode".
Calibrate pH sensor
✱
Perform calibration, see chapter 8 "Calibrating a pH measurement chain",
page 51.
Calibrate chlorine sensor
✱
The process must now be brought to the state that is as relevant as
possible to the final value (e.g. when measuring chlorine, the required
concentration).
✱
Start the calibration, see "Ways to start the calibration", page 71.
✱
Select the limit point calibration with the
key.
✱
Wait until the display value has stabilized; then press
to continue.
Set the displayed value to the measured reference value with the
or
keys; then press
to continue.
The pH signal and temperature signal are supplied via the main input, the
chlorine signal (standard signal) via the optional input.
PGM
PGM
PGM
Summary of Contents for dTRANS pH 02
Page 66: ...9 Calibrating a redox measurement chain 66...
Page 89: ...89 12 Calibration logbook Next most recent successful calibration Briefly press the key...
Page 100: ...14 Setup program 100 Export data for processing in an external program...
Page 133: ......
Page 134: ......
Page 135: ......