11 Calibrating a sensor with a standard signal
78
11.4 Conductivity operating mode
11.4.1 Calibration of the relative cell constant
The transmitter is in "Measuring mode".
✱
Immerse the conductivity sensor in a reference solution with a known
conductivity.
✱
Start the calibration, see "Ways to start the calibration", page 71.
✱
Select REL. CELL CONST.
✱
Press the
key.
✱
When the measured value is stable, press the
key
✱
The measured conductivity value flashes on the display.
✱
Use the
or
keys to set the value to the actual conductivity.
This example is based on a conductivity sensor with a connected two-wire
transmitter.
PGM
PGM