11 Calibrating a sensor with a standard signal
72
11.2 Linear operating mode
11.2.1 1-point calibration
The transmitter is in "Measuring mode".
✱
Now bring the system to a defined state (e.g. when measuring liquid level,
empty the container).
✱
Start the calibration, see "Ways to start the calibration", page 71.
✱
Select the zero point calibration with the
key.
✱
Wait until the display value has stabilized; then press
to continue.
Set the displayed value to the required value (usually 0%) with the
and
keys; then press
to continue.
The zero point determined by the device is displayed.
This example is based on a liquid level measurement (as a %).
The input signal is provided by a pressure transmitter.
PGM
PGM
PGM