9 Calibrating a redox measurement chain
62
9.3
Zero-point calibration (one-point offset calibration)
✱
Make preparations, see chapter 9.2 "General information", page 60.
✱
Start calibration, see chapter 9.2.2 "Ways to start the calibration", page 61.
✱
Select zero point calibration.
✱
Immerse the combination electrode in a test solution with a known redox
potential.
✱
Start the zero point calibration with the
key.
Wait until the display value has stabilized; then press
to continue.
✱
Set the displayed value to the test solution value with the
or
keys;
then press
to continue.
The zero point determined by the device is displayed.
✱
Use the
key to accept the value or
the
key to reject it.
Zero point calibration is only available if the unit is configured as "mV"!
PGM
PGM
PGM
PGM
EXIT