59
8 Calibrating a pH measurement chain
keys; then press
to continue.
The zero point of the combination electrode determined by the device and its
slope in the acidic and alkaline ranges of the characteristic curve are also
displayed.
✱
Use the
key to accept the calibrated values or
reject them with the
key.
The device returns to measuring mode.
8.6
pH Antimony measurement chains, ISFET pH
combination electrodes
Antimony measurement chains and ISFET pH combination electrodes are
calibrated similarly to "normal" pH measurement chains.
- General information on calibration see "General information", page 51.
- Zero point calibration see chapter 8.3 "Zero point (1-point) calibration",
- 2-point calibration see chapter 8.4 "2-point calibration", page 54.
- 3-point calibration see chapter 8.5 "3-point calibration", page 57.
PGM
PGM
EXIT
IIf the following permissible limits of the calibration values are not observed in
the calibration procedure then an error is displayed at the end of the proce-
dure:
Antimony electrode:
-2 ... 2 pH, slope 10 ... 110 %
Standard glass electrode
5 ... 9 pH, slope 75 ... 110 %