5 Installation
22
5.4.4
Symmetrical connection of a combination electrode with separa-
te temperature sensor
✱
Connect the core wires according to the terminal assignment; see below
and see chapter 5.3 "Connection", page 15.
For notes on the application, see "Symmetrical connection of pH electrodes",
page 117.
(1)
pH combination electrode
(2)
Ground pin or conductive pipe/container wall at the measuring point
(3)
Separate temperature sensor
(4)
Double shielded coaxial cables (triaxial cables)
?
1
2
3
4
1
2
3
4
5
6
7
8
6
7
8
9
10
9
10
11
12
4
L1(L+)
5
6
8
9
11
12
13
15
16
17
N(L+)
11
12
13
14
15
(3)
(2)
(1)
(4)
The premounted bridge (8-9) must be removed!
Double shielded coaxial cables (triaxial cables) must be used in environments
with difficult EMC conditions. A shielded 2-core cable is required to connect a
temperature probe.