Display of measured values STANDARD
The measurement value, measurement variable and temperature of the
measuring material are shown in standard display.
Display of measured values TENDENCY
The operator can quickly see the direction in which the measurement is
changing.
AF8
HySt
w
x
0
1
(1)
(2)
(3)
Operating mode
Display bottom (temperature input)
Display top (analog input measurement value)
(3)
(1)
(2)
Rising
Steady
Falling
Greatly
Moderately
Slightly
Slightly
Moderately
Greatly
The measurement tendency (trend) is calculated over the last 10 measurement
values.
So with a sampling interval of 500 ms, the last 5 seconds are considered.