9 Calibrating a sensor with a standard signal
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
9.4.2
Calibration of the temperature coefficient
Linear temperature coefficient
The transmitter is in "Measuring mode".
✱
Immerse the conductivity sensor in the sample medium.
Start the calibration, see "Ways to start the calibration", page 54.
✱
Select "LINEAR TEMP. COEF.".
The current sensor temperature flashes in the display (1).
✱
Enter the required working temperature and confirm your entry.
This example is based on a conductivity sensor with a connected two-wire
transmitter.
(1)
The working temperature must be at least 5 °C above or below the reference
temperature (25.0 °C).