
59
9 Calibrating a sensor with a standard signal
✱
Use the
key to accept the value or the
key to reject it.
✱
The instrument returns to Measuring mode.
Calibration is complete
After rinsing, the sensor can again be used to take measurements.
9.3
pH operating mode
9.3.1
Zero-point (1-point) calibration
The transmitter is in "Measuring mode".
✱
Perform calibration as follows.
Zero point (1-point) calibration
✱
Make preparations, see chapter 8.2 "General information", page 47 .
✱
Start calibration, see chapter 8.2.3 "Ways to start the calibration", page 48.
PGM
EXIT
This example is based on a glass combination electrode with a connected
two-wire transmitter.
Summary of Contents for dTRANS CR 02
Page 83: ...83 12 Setup program Export data for processing in an external program...
Page 118: ...17 Index 118...
Page 119: ......