Chapter 4 _________________________________________________________________ Operation
JUMO __________________________________________________________________________ 99
Data Filtering
The averaging data filter calculates an average over a certain period of
time. The lowest measurement noise is achieved with the extended
filtering. There are three filtering levels available.
Table 26
Filtering Levels
Setting
Filtering level
OFF
No filtering
ON (default)
Standard = short filtering (approximately 15 s moving
average)
EXTENDED
Extended filtering (default: approximately 1 min
average)
Use display/keypad to set the filtering level.
1.
Press any of the arrow buttons to open the
MAIN MENU
.
2.
Select
Measuring
by pressing the ► arrow button.
3.
Select
Filtering
and press
CHANGE
to confirm your selection.
4.
Select
Off/Standard/Extended
and press
SELECT
to confirm
your selection.
5.
Press
EXIT
to return to the basic display.
FILT
Use the serial line command
FILT
[
xxx
] to set the filtering level.
FILT
[
xxx
]
where
xxx =
OFF, ON or EXT (default = OFF)
Device Information
Use the display/keypad or the serial line to display the device
information.
Press the
INFO
button in the basic display to see the following
information: