16 Calibrating Ci conductivity sensors
80
TC curve (for nonlinear temperature coefficients)
If the conductivity of a liquid whose temperature coefficient changes with tem-
perature has to be measured, this method can determine 5 temperature coeffi-
cients for 5 temperature intervals. In this way, it is possible to determine a good
approximation of the temperature coefficient curve. While the operator brings
the sample solution to the temperature values requested by the device, the de-
vice determines the temperature coefficient for each interval. This requires in-
stallation of a temperature sensor that the device can use to sense the
temperature of the sample solution.
16.2.2
Calibration presets for Ci conductivity sensors
Before you can perform a calibration, you must first enter the necessary calibra-
tion default settings. The possible settings for the Ci calibration are described in
the following.
Open the calibration default settings:
Device menu > Calibration > Select Ci Analysis Input or Universal Input >
Calibration Presets
Conductivity
6 calibration points
5 intervals
Temperature
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
⇨
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......