![JUMO AQUIS touch P 202580 Operating Manual Download Page 28](http://html1.mh-extra.com/html/jumo/aquis-touch-p-202580/aquis-touch-p-202580_operating-manual_2031168028.webp)
5 Mounting
28
5.2
Dimensions
Spacings
For adequate ventilation, the following mounting distances from adjacent equip-
ment must be respected:
•
horizontal 35 mm
•
vertical 80 mm
When inserting the device, sufficient space for maneuvering the connecting ca-
bles must be left behind the case. A Ci-conductivity sensor is connected via a
M12 connector adapter. This increases the space requirement with respect to
mounting depth (see dimensional drawing below).
92 mm
+0.8 mm
92 mm
+0.8 mm
130,9 mm
96 mm
123,1 mm
90,5 mm
97,1 mm
104,1 mm
96 mm
22,25 mm
22,25 mm
22,25 mm
22,25 mm
90,5 mm
97,1 mm
104,1 mm
5,5 mm
Panel thickness max. 5 mm
Summary of Contents for AQUIS touch P 202580
Page 12: ...Content...
Page 26: ...4 Identifying the device version 26...
Page 29: ...29 5 Mounting Mounting depth with Ci conductivity sensor 283 3 mm 2 3 mm 36 mm 25 mm 23 2 mm...
Page 32: ...5 Mounting 32...
Page 64: ...7 Startup 64...
Page 100: ...1 2 3 4 5...
Page 107: ...107 8 Operation Example of a Quad View...
Page 108: ...8 Operation 108...
Page 114: ...9 Parameterization 114...
Page 160: ...10 Configuration 160...
Page 182: ...13 Calibrating a pH measuring chain 182...
Page 196: ...16 Calibrating CR conductivity sensors 196...
Page 206: ...17 Calibrating Ci conductivity sensors 206...
Page 214: ...18 Calibrating universal inputs 214...
Page 228: ...19 PC Setup Program 228...
Page 319: ...319 20 Technical data...
Page 320: ...20 Technical data 320...
Page 321: ......