6 Operation
38
✱
ADMINISTRATOR LEVEL > PASSWORD > PARAMETER LEVEL > CTRL.
SPEC. FUNCT. > MANUAL MODE: set LOCKED or PULSED or SWITCH-
ING.
LOCKED=
No manual mode; the JUMO AQUIS 500 is in control
PULSED
The outputs are active for as long as the
or
key is
pressed
SWITCHING
The outputs become active when the
or
key is
pressed; when the relevant key is pressed again, the corre-
sponding output is deactivated again
Activating manual mode
The device is in measurement mode.
✱
Press the
and
keys for less than 2 seconds.
"MAN." appears in the status line of the LCD display.
The device outputs then behave according to the pre-settings.
The JUMO AQUIS 500 is no longer in control. The output level at the output of
the controller channel is 0 %.
To select controller channel 1, use the
key; the output level at the output of
the controller channel 1 is set to 100 %.
To select controller channel 2, use the
key; the output level at the output of
controller channel 2 is set to 100 %.
MANUAL/simulation overview
It is possible to display which outputs or controllers are in MANUAL mode.
Requirement: the device is in measurement mode.
✱
Press the
key several times for less than 2 seconds (number varies
according to the features and configuration of the device).
NOTE!
If the
and
keys are pressed for more than 3 seconds, the device enters
HOLD mode.
To exit HOLD mode, press the
and
keys again for more than 3 seconds.
Summary of Contents for AQUIS 500 RS
Page 5: ...Contents 12 5 Template for panel cut out 114 13 China RoHS 116 14 Index 117...
Page 8: ...2 Description 8 Block diagram...
Page 16: ...4 Mounting 16...
Page 112: ...12 Appendix 112 12 4 2 Manual configuration of a sensor 1 Only for temperature sensor MODBUS...
Page 113: ...113 12 Appendix...
Page 114: ...12 Appendix 114 12 5 Template for panel cut out...
Page 115: ...115 12 Appendix...
Page 119: ......