17 Calibrating universal inputs
17.3.3
Slope calibration (free chlorine, ph/Temp.-compensated)
Step
Action
1
Start the slope calibration.
Device Menu
r
Calibration
r
Select Universal Input
r
Slope Cal-
ibration
2
Clean the sensor and immerse it in the test solution.
3
Check the displayed values for the pH value and
temperature. Automatic sensing can be configured in the calibra-
tion default settings for both process variables independently of
one another. With automatic sensing, the particular influencing
variable is only displayed and can no longer be changed here.
Without automatic sensing, the particular influencing variable must
be entered manually here.
Wait until the measuring value displayed stabilizes and then check
the pH value displayed. Then confirm by pressing "OK".
Example with temperature sensing and without pH-value sensing
4
Enter the concentration value of the test solution. Confirm by
pressing "OK"
5
A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
Tap button to enter pH
value of test solution
Measuring value
based on previous
calibration values
Tap button to enter
chlorine concentration
Summary of Contents for 202580
Page 6: ...Content...
Page 28: ...12 Calibrating a pH measuring chain 28...
Page 42: ...15 Calibrating CR conductivity sensors 42...
Page 52: ...16 Calibrating Ci conductivity sensors 52...
Page 60: ...17 Calibrating universal inputs 60...
Page 79: ...79 18 Technical data...
Page 80: ...18 Technical data 80...
Page 81: ......